Detector types are selected depending on the particular measurement application , the required detection limit, matrix interferences, and/or regulatory guidelines. Since all five of our chassis can mount up to four detectors simultaneously (some can mount five or even six), you can perform a surprising variety of applications with one instrument.
DETECTOR SELECTIVITY SENSITIVITY (approximate)
CCD hydrogen and hydrocarbons 500ppm
TCD universal 200-500ppm
FID hydrocarbons 1ppm
DELCD chlorinated and brominated molecules 10ppb
FID/DELCD hydrocarbons, chlorinated and brominated molecules 10ppm
HID universal, except neon 10ppm
PID aromatics and molecules with double carbon bonds 10ppb
NPD nitrogen and phosphorus 10ppm
NPD/DELCD nitrogen, phosphorus, chlorinated and brominated molecules 100ppb
TID nitro functional groups (TNT, etc.; explosives) 1ppb, 50ppb
ECD electronegative compounds 10ppb
RGD reducing gases (like hydrogen and CO) 10ppb
FPD sulfur or phosphorus 200ppb and 10ppb
FPD/FID sulfur or phosphorus, and hydrocarbons 200ppb, 10ppb, and 100ppm
Dual FPD sulfur and phosphorus simultaneously 200ppb and 10ppb
FID Dual FPD hydrocarbons, sulfur and phosphorus simultaneously 100ppm, 200ppb, and 10ppb