Injector types are selected by the user depending on the particular measurement application, detection limit, and regulatory requirements. Each GC chassis except the 110 comes standard with an On-Column injector. The Heated Flash Vaporization, Heated Split/Splitless, and PTV injectors are upgrades to the standard On-Column injector. There are 17 injector types from which to choose. Up to five injectors may be mounted simultaneously on Models 8610C and 8610D. The Model 310 will accommodate a single On-Column, Heated Flash Vaporization, Heated Split/Splitless, or PTV injector. Rack-mountable Model 410 can accommodate a gas sampling valve and a single On-Column, Heated Flash Vaporization, Heated Split/Splitless, or PTV injector.Sample Types and Appropriate Injectors
LIQUIDS: On-column, heated flash vaporization, heated split/splitless, PTV, heated static headspace, purge & trap, liquid injection valve, or liquid autosampler
SOLIDS: Thermal desorber, heated static headspace, PTV, or Method 5035 purge & trap
GASES: On-column, gas sampling valve, Method TO-14 air concentrator, or heated static headspace
SPME FIBERS: Heated flash vaporization with low volume SPME liner, or heated split/splitless